Rockwell Automation 1440-DYN02-01RJ XM Dynamic Measurement Module Manuel d'utilisation

Page 69

Advertising
background image

Publication ICM-UM002F-EN-E - March 2013

69

Configure XM Module in RSLogix 5000 Software

Chapter 3

2.

Configure the parameters, as necessary.

In this field

Values are

Comments

Enable Auto Trigger

Check to enable Auto Trigger mode. The minimum
signal amplitude for triggering is 2 volts peak-to-peak
and the minimum frequency is 6 CPM (0.1 Hz).

Clear to enable Manual Trigger mode. The value
entered in Trigger Threshold is used as the trigger
point. The minimum signal amplitude for triggering is
500 millivolts peak-to-peak and the minimum
frequency is 1 CPM (0.016 Hz).

Using Auto Trigger mode may cause the
tachometer to trigger on noise if the signal is
very small. For example, you have 1 volt of
noise on a 2 volt signal. To prevent this, make
sure the % noise in the signal is less than the
value entered in the Trigger Hysteresis.

Trigger Hysteresis

Enter the amount of hysteresis around the trigger
threshold.

Enter a value between 0...50.

In Auto Trigger mode, the value entered is a
percentage of the peak-to-peak input signal.

In Manual Trigger mode, the value entered is a
voltage level. The hysteresis voltage is added
to or subtracted from the threshold voltage to
determine the hysteresis range.

Trigger Level

Enter the signal level to be used as the trigger value
when in Manual Trigger mode.

This parameter is dimmed in Auto Trigger
mode.

Trigger Slope

Choose the input signal slope to be used with the
trigger value.

• Positive
• Negative

The trigger point of the tachometer defines 0° for
phase measurement. If the tachometer is a square
wave, the phase angles measured will vary by 180°
depending on whether the Trigger Slope is set to
positive or negative.

DC High Limit

Enter the maximum expected DC bias voltage from the
transducer.

A voltage reading outside this range
constitutes a transducer fault, which is
indicated with the tachometer LED blinking red
and the TachFault input tag.

For information about the LEDs,

see page 87

.

For information about the input tags,

see

page 91

.

DC Low Limit

Enter the minimum, or most negative, expected DC
voltage from the transducer.

Inhibit Zero Pulse Tachometer
Fault

Check to enable Inhibit Zero Pulse Tachometer Fault.

Clear to disable Inhibit Zero Pulse Tachometer Fault.

Controls whether a tachometer fault occurs if
no pulses are detected on the tachometer
signal.

Fault Delay

Enter the number of seconds that the module should
wait after the last valid pulse signal before it indicates
a tachometer fault.

Enter a value between 1...64 seconds.

Advertising