Metrohm 730 Sample Changer Manuel d'utilisation

Page 99

Advertising
background image

4.4

Changer Commands

730 Sample Changer, Detailed Description

96

PUMP

PUMP

4

>sample sequence

4 PUMP 1.1

: 1 s

1.1…2.2

1…999 s,

1.

,2.

ON,OFF

Pump control

Up to 4 pumps (2 pumps/tower) can be separately controlled with
the PUMP command. The choice of pump is indicated in the first
parameter.
The syntax: T.P (T=Tower number, P=Pump)
Only 2 pumps can be in operation simultaneously. Pump 1 of each
tower does the rinsing of the titration head. Pump 2 (if present)
can be used for removing of the sample solution by aspiration.
Combined rinsing and aspiration is possible with 'PUMP 1.

' or

'PUMP 2.

' .

The pumps can be turned off and on as desired or operated for a
specific amount of time. The LEARN function is useful for determi-
ning the interval of time for optimal rinsing or aspiration. See also
page 50.

STIR

5

STIR

>sample sequence

5 STIR 1

: 1 s

1…4,

ON,OFF,

1…9999 s

Stirrer control

Up to 4 stirrers can be individually controlled with the STIR com-
mand. The choice of stirrer is indicated in the first parameter. With
'STIR

' all stirrers can be operated simultaneously.

The stirrers can be specifically switched on or off or operated for a
certain time interval.
In the Parameter Menu under '>stirring rates' the speed of each
individual stirrer can be determined specific to a method. This can
also be modified in a sequence using the corresponding 'DEF'
command.

DOS

DOS

6

>sample sequence

6 DOS 1 : 1 ml

1…12,

fill,

release, prepar.,

empty, eject,
adjust, level

0.001…1…999.999 ml

Dosing Control