4.5

Sample Racks

730 Sample Changer, Detailed Description

108

Sample method for the use of special beakers:

Electrode calibration in a start sequence

LIFT: 1 :

shift mm

MOVE 1 :

spec.1

Spec.1 = rinsing beaker

LIFT: 1 :

rinse mm

PUMP 1.1 :

3 s

rinse electrode

WAIT

2 s

LIFT: 1 :

shift mm

MOVE 1 :

spec.2

spec.2 = buffer 1

LIFT: 1 :

work mm

STIR: 1 :

on s

CTL:Rm: START device1

start measuring buffer 1

SCN:Rm:

endmeter 1

measurement finished? then ...

STIR: 1 :

off s

LIFT: 1 :

shift mm

MOVE 1 :

spec.1

spec.1 = rinsing beaker

LIFT: 1 :

rinse mm

PUMP 1.1 :

3 s

rinse electrode

WAIT

2 s

LIFT: 1 :

shift mm

MOVE 1 :

spec.3

spec.3 = buffer 2

LIFT: 1 :

work mm

STIR: 1 :

on s

CTL:Rm: START device1

start measuring buffer 2

SCN:Rm:

endmet 1

measurement finished? then...

STIR: 1 :

off s

...