8 remote interface, 8 remote interface remote interface – Metrohm 730 Sample Changer Manuel d'utilisation

Page 122

Advertising
background image

4.8 Remote Interface

730 Sample Changer, Detailed Description

119

4.8 Remote Interface

Remote Interface

Peripheral instruments connected such as Titrinos, Titroproces-
sors, pH Meters etc. can be controlled via the remote interface
(25-pin socket).
14 lines (Output 0–13) are available for the emission of signals.
For receiving signals (e.g. the "ready" signal of a Titrino at the end
of a titration) 8 lines are provided (Input 0–7).

Pin Assignment of the Remote socket:

Output 6
Output 7
Output 4
Output 2
Output 0
Output 8
Output 9
Output 10
Input 1
Input 3
Input 5
Input 7
Output 11

0 Vo lt

+5 Volt

Output 5
Output 3
Output 1

Output 12
Output 13

Input 0
Input 2
Input 4
Input 6

0 Vo lt

The +5 V supply line may by charged with 20 mA maximally.

When the 759 Swing Head is connected to the remote interface,
the output lines 11–13 and the input line 7 are occupied. These
four lines are not continued in the plug and ignored, when further
instruments are connected via the remote cable (see page 15ff).

The coupling with Metrohm instruments with the standard cable
6.2141.020

is normally resolved in the following manner:

730

Metrohm instrument

³

730

Metrohm instrument

³

Output 0

–––––––

Input 0

³

Input 0 –––––––

Output 0

Output 1

–––––––

Input 1

³

Input 1 –––––––

Output 1

Output 2

–––––––

Input 2

³

Input 2 –––––––

Output 2

Output 3

–––––––

Input 3

³

Input 3 –––––––

Output 3

Output 4

–––––––

Input 4

³

Input 4 –––––––

Output 4

Output 5

–––––––

Input 5

³

Input 5 –––––––

Output 5

Output 6

–––––––

Input 6

³

Input 6 –––––––

Output 6

Output 7

–––––––

Input 7

³

Input 7 –––––––

Output 7

Output 8

–––––––

free

Output 9

–––––––

"

Output 10 –––––––

"

Output 11 –––––––

"

Output 12 –––––––

"

Output 13 –––––––

"

+5V

active = low

inactive = high

active = low

inactive = high

Inputs

Outputs

The Output lines 8…13 are unused by

other Metrohm instruments as yet.

They are assigned to pin 6…8, 13,

19…20.

Advertising