5.4 Diagnosis

730 Sample Changer, Appendix

171

The key code table:

5.4.6 Remote Interface

This diagnosis step tests the functionality of all the output (14) and
input lines (8).

Prepare instrument for diagnosis (see chapter 5.4.2).

If necessary, press <ê> several times until

diagnosis

>remote test

<ENTER>

>remote test

remote test connector ?

Plug test plug 3.496.8550 into the remote interface without
turning the instrument off.

<ENTER>

The test runs automatically. If no error occurs, the following appears:

>remote test

remote test ok

Remove test plug and press <ENTER>.

diagnosis

>RS232 test

Code

Key

Code

Key

1

<HOLD / LEARN>

16

<7 / SAMPLE>

2

<STOP>

17

<4 / PUMP>

3

<START

18

<1 / SCAN>

4

<CONFIG>

19

<0 / DEF>

5

<PARAM>

20

<END>

6

<USER METHOD>

21

<è>

7

22

<CLEAR / RESET>

8

<9 / LIFT>

23

<ENTER>

9

<6 / DOS>

24

<é>

10

<3 / WAIT>

25

<ê>

11

<* / ENDSEQ>

26

<SELECT / TOWER>

12

<8 / MOVE>

27

<QUIT>

13

<5 / STIR>

28

<HOME>

14

<2 / CTRL>

29

<ç>

15

<. / PRINT>

30

<INSERT >

31

<DELETE>