background image

4.7 Pipetting with the Swing Head

730 Sample Changer, Detailed Description

117

Configuration

Dosino 1:

Pipetting Dosino

Dosino 2:

For dosing auxiliary solution 1

Dosino 3:

For dosing auxiliary solution 2

Sample rack 6.2041.400 for 126 x 15 mL and 2 x 250 mL vessels

Lift positions:

Working position:

Tower 2, lift height for titration

Rinsing position:

Tower 1, lift height for rinsing pipette tip

Shift position:

Tower 1 and Tower 2, lift position for shifting
the sample rack

Special position:

Tower 1, pipette tip submerged in sample to
aspirate the solution

Lift position 5:

Tower 1, pipette tip above the sample solution

Lift position 6:

Tower 2, pipette tip submerged in titration
solution to eject the sample solution

Lift position 7:

Pipette tip on scraper

Special beaker positions: Spec 1: Titration vessel

Spec 2: Rinsing vessel
Spec 3: Scraper

To scrape away any drops of liquid remaining on the pipette tip after rinsing the pipet-
ting tube, a "scraper" needs to be installed. This can be e.g. a test tube with a 1 mL
pipette tip for air cushion pipettes fixed at its center. The position of the "scraper" is
defined as special beaker 3.

730 Sample Changer 730.0013
Parameters
method PipMeth1
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 MOVE 1 : spec.2
3 LIFT: 1 : rinse mm
4 DRIVE.PORT 1.4: dos.
5 DOS: 1 : eject ml
6 LIFT: 1 : shift mm
>sample sequence
1 MOVE 2 : spec.1
2 LIFT: 2 : work mm
3 DOS: 2 : 100 ml
4 DOS: 3 : 2 ml
5 STIR: 1 : 5 s
6 LIFT: 2 : shift mm
7 MOVE 1 : spec.2
8 LIFT: 1 : rinse mm
9 DOSRATE 1 15
10 FILLRATE 1 15
11 DOS: 1 : fill ml
12 DRIVE.PORT 1.4: dos.
13 DOS: 1 : 5 ml
14 DRIVE.PORT 1.1: dos.
15 DOS: 1 : eject ml
16 DOSRATE 1 5
17 FILLRATE 1 5

The start sequence consists of commands which prepare

the system for pipetting.

To remove air bubbles from the system, the content of the

cylinder is ejected via port 4.

Titration vessel in front of tower 2

Auxiliary solution 1 added to the titration vessel

Auxiliary solution 2 added to the titration vessel

Rinsing station in front of tower 1

Lift 1 rises to rinsing height

Dosing and filling rates for Dosino 1

Filling of cylinder

Cock opened to port 4

Ejection of 5 mL of water via port 4

Cock opened to port 1

Content of cylinder ejected via pipetting tube

Dosing and filling rates for Dosino 1

Advertising