Metrohm 730 Sample Changer Manuel d'utilisation

Page 21

Advertising
background image

2.5 Integration

730 Sample Changer, Installation

18

Remote control commands:

CTL:Rm : START device1

starts Titrino 1

CTL:Rm : *************1

"

CTL:Rm : START device2

starts Titrino 2

CTL:Rm : ********1*****

"

CTL:Rm : START device*

starts both Titrinos at the same time

CTL:Rm : ********1****1

"

Scan commands:

SCN:Rm : end1

awaits end of titration of Titrino 1

SCN:Rm : ****1***

"

SCN:Rm : end2

awaits end of titration of Titrino 2

SCN:Rm : *1******

"

SCN:Rm : ready1

awaits readiness of Titrino 1

SCN:Rm : *******1

"

SCN:Rm : ready2

awaits readiness of Titrino 2

SCN:Rm : **1*****

"

SCN:Rm : ready*

awaits readiness of both Titrinos

SCN:Rm : **1****1

"

Operating Titrinos and pH meters at the same time is also possi-
ble in the same manner.

730

B

A

Titrino

C

D

692
712
713

device 1

device 2

Cable 6.2141.030

Advertising