Metrohm 730 Sample Changer Manuel d'utilisation
Page 156

4.10 Remote Control Commands
730 Sample Changer, Detailed Description
153
4.10.2.58
A
ssembly.Def
$G
A
ssembly.Def.Object
STIRRATE
, DOSRATE
FILLRATE, LIFTRATE
SHIFTRATE, DRIVE.PORT
A
ssembly.Def.Adress
depending on object
A
ssembly.Def.Value
depending on object
Definition of various device settings. Depending on the selected
item in …DEF.Object different parameters and ranges may be
entered.
Def.Object
Def.Address
Def.Value
STIRRATE
1
…4
1…3…15
DOSRATE
1
…12
0.01…160 mL/min
FILLRATE
1
…12
0.01…160 mL/min
LIFTRATE
1
…2
3…25 mm/s
SHIFTRATE
auto.
, +, –
3…20 ang. degrees/s
DRIVE.PORT
[1…12].[1…4]
dos.,
fill, rinse, prep., drain
&Assembly.Def;$G triggers this function.
4.10.2.59
A
ssembly.Wait
$G, $S
A
ssembly.Wait.Time
0...1...9999 s
Waiting time.
&Assembly.Wait;$G triggers this function.
4.10.2.60
A
ssembly.End
$G
$G triggers a RESET (same as the <ENDSEQ> key).