background image

4.10 Remote Control Commands

730 Sample Changer, Detailed Description

153

4.10.2.58

A

ssembly.Def

$G

A

ssembly.Def.Object

STIRRATE

, DOSRATE

FILLRATE, LIFTRATE

SHIFTRATE, DRIVE.PORT

A

ssembly.Def.Adress

depending on object

A

ssembly.Def.Value

depending on object

Definition of various device settings. Depending on the selected
item in …DEF.Object different parameters and ranges may be
entered.

Def.Object

Def.Address

Def.Value

STIRRATE

1

…4

1…3…15

DOSRATE

1

…12

0.01…160 mL/min

FILLRATE

1

…12

0.01…160 mL/min

LIFTRATE

1

…2

3…25 mm/s

SHIFTRATE

auto.

, +, –

3…20 ang. degrees/s

DRIVE.PORT

[1…12].[1…4]

dos.,

fill, rinse, prep., drain

&Assembly.Def;$G triggers this function.

4.10.2.59

A

ssembly.Wait

$G, $S

A

ssembly.Wait.Time

0...1...9999 s

Waiting time.
&Assembly.Wait;$G triggers this function.

4.10.2.60

A

ssembly.End

$G

$G triggers a RESET (same as the <ENDSEQ> key).