background image

4.9 Operation via RS232 Interface

730 Sample Changer, Detailed Description

128

The following global status conditions are possible:

$G

G

o

The sample changer is executing the last command.

$H

H

old

The sample changer has been held ($H, <HOLD> key or by an
error which effects the hold status).

$C

C

ontinue

The sample changer has been restarted actively after hold.

$R

R

eady

The sample changer has executed the last command and is
ready.

$S

S

top

A process has been aborted, e.g. by pressing the <STOP> key
or because there was an error.

Detailed Status Conditions

Status conditions of the global $R:

$R.Mode

Basic state: ready to start automatic processing

$R.Assembly

An assembly step has been executed.

Status conditions of the global $G:

$G.Mode.Start.

Instrument at the beginning of processing

$G.Mode.Start.01.WAIT

Instrument processing the start sequence, displays line number
and current command

$G.Mode.Sample.01.WAIT

Instrument processing the sample sequence, displays line
number and current command

$G.Mode.Final.01.WAIT

Instrument processing the final sequence, displays line number

and current command

$G.Mode.

Instrument processing a manual command

$G.Assembly.

Instrument processing an assembly-command

Status conditions of the global $H:

$H.Mode

The status conditions of the global $H are identical with the
ones of the global $G.

4.9.5 Error Messages, Errors

Fatal Instrument Errors:

E1

Incorrect program check sum

E2

RAM read/write error

E3

RAM lost data

E4

Timer interrupt for multi-tasking missing

E5

RS232 module test error

E6

RS232 read/write error

E7

Display read/write error

E12

EBUS error

E18

Low battery

E19

RAM test error