4.2 The Keyboard

730 Sample Changer, Detailed Description

73

Key

Normal State and Editing

DEF commands are valid for manual operation as well as for
the programmed processing of a method.

The individual DEF commands are listed below.

0

DEF

STIRRATE

Change stirring speed

Change stirring speed

The stirring speed can be individually regulated for every
stirrer (rod or magnet stirrer).

Syntax: STIRRATE [Stirrer-No.] [Stirring speed]

0

DEF

DOSRATE

Change dosing rate

Change dosing rate

The dosing rate may be individually set for every dosing
drive (Dosimat or Dosino).

Syntax: DOSRATE [Dosing unit] [Dosing rate]

0

DEF

FILLRATE

Change filling rate

filling rate

The filling rate may be set individually for every dosing drive
(Dosimat or Dosino).

Syntax: FILLRATE [Dosing unit] [Filling speed]

0

DEF

LIFTRATE

Change lift speed

The lift speed can be set for both towers (if two are present).

Syntax: LIFTRATE [Tower] [Lift speed]

0

DEF

SHIFTRATE

Change turning speed and

turning speed and direction

In addition to giving the turning speed of the sample rack in
angular degrees/sec., the turning direction can also be indi-
cated with the first parameter.

Turning direction "+" causes the sample beakers to be proc-
essed in the counterclockwise direction, i.e. in ascending
order. Turning direction "–" indicates in the clockwise direc-
tion, i.e. in descending order. The beaker positions are
numbered on every sample rack and easily visible.

With turning direction "auto" the sample changer independ-
ently chooses the shortest possible path for placing a
beaker under a particular lift. The turning direction is auto-
matically chosen.

Syntax: SHIFTRATE [Turning direction] [Turning speed]