2.2 Power Supply

730 Sample Changer, Installation

8

Disconnect line cable
Unplug the 730 Sample Changer.

Remove fuse holder
Using a screw driver, loosen the fuse holder and pull it out.

Checking and replacing fuse
Carefully remove the built-in fuse and check its specifications.
(The position of the fuse in the fuse holder is marked by the
white arrow printed next to the supply voltage):

100

…120 V 0.5 A (slow) ord. no. U.600.0014

220

…240 V 0.25 A (slow) ord. no. U.600.0011

Replace fuse
Replace fuse if necessary and reinsert it in the fuse holder.

Insert the fuse holder
Insert the fuse holder according to the appropriate supply volt-
age. The white arrow besides the desired voltage has to point
towards the white block mark printed on the fuse holder's panel
(see below).

100 – 120 V

220 – 240 V

Power switch

Power plug

Fuse holder