5.4 Diagnosis

730 Sample Changer, Appendix

168

Required Instruments:

Only required, if the RS232 or remote interface is to be checked:
Test plug 3.496.8550 (on the 'Remote' plug)
Test plug 3.496.8480 (on the 'RS 232' plug)

5.4.2 Preparing the Instrument

Disconnect power plug

Remove cables to the RS232 and remote interfaces.

Plug in power plug and immediately press <9> and keep it de-
pressed until the start test pattern disappears.

Alternatively, the <CLEAR> key can be pressed to cause a re-
set followed by pressing (within 0.4 sec) and holding down the
<9> key to open the diagnosis menu.

Switch on and

9

LIFT

or

RESET

CLEAR

and

9

LIFT

Main Menu Diagnosis:

diagnosis

>RAM initialization

Open the submenu with <ENTER>

diagnosis

>RAM test

Use <

é

> or <ê> to access one menu item up or

down

diagnosis

>display test

Use <HOME> or <END> to access first or last
menu item

diagnosis

>display contrast test

<QUIT> returns to the normal state

diagnosis

>key test
diagnosis

>remote test
diagnosis

>RS232 test
diagnosis

>external bus test
diagnosis

>beeper test
diagnosis

>rack code test
diagnosis

power on reset

The "function test" is a Metrohm internal test.