Changer Commands
730 Sample Changer, Detailed Description
102
DEF
0
DEF
Redefining specific instrument settings
During method processing the most diverse settings can be made
using the following DEF commands. The individual entries are
accessed by repeatedly pressing the DEF key (roll-up selection).
>sample sequence
13 STIRRATE
1
3
1…4
1…3…15
Stirring speed
The stirring speeds for all 4 stirrers can be individually set. The
first parameter stands for the number of the stirrer; the second pa-
rameter allows the setting of the stirrer speed in 15 steps.
>sample sequence
14 DOSRATE 1
160
1…12
0.01…160 ml/min
Dosing speed
The dosing speed for all 12 dosing units can be individually set.
The first parameter stands for the number of the dosing unit and
the second parameter allows setting the dosing speed in mL/min.
>sample sequence
15 FILLRATE 1
160
1…12
0.01…160 ml/min
Filling speed
The filling speed can be set for all 12 dosing units individually. The
first parameter stands for the number of the dosing unit and the
second parameter allows setting the filling speed in mL/min.
>sample sequence
16 LIFTRATE 1
25
1,2
3…25 mm/s
Lift rate
The lift rate can be set for both towers individually. The first para-
meter stands for the number of the tower and the second para-
meter allows setting the lift rate in mm/s.