4.10 Remote Control Commands

730 Sample Changer, Detailed Description

148

4.10.2.40

S

etup.Lock.Keyboard

on, off

S

etup.Lock.Config

on, off

S

etup.Lock.Parameter

on, off

S

etup.Lock.UserMeth.Recall

on, off

S

etup.Lock.UserMeth.Store

on, off

S

etup.Lock.UserMeth.Delete

on, off

S

etup.Lock.Display

on, off

"on" means disable the corresponding function.
.Keyboard

Disables all keys of the keyboard, except the
<START>, STOP> and <CLEAR> key.

.Config

Locks the configuration menu

.Parameter

Locks the parameter menu

.Usermeth.Recall Locks the function "recall method"
.UserMeth.Store Locks the function "store method"
.Usermeth.Delete Locks the function "delete method"
.Display

Disables the LCD display. The Sample
Changer will not support the display.

4.10.2.41

S

etup.Mode.StartWait

on, off

Indefinite start delay. Only for remote control.

4.10.2.42

S

etup.AutoInfo.Status

on

, off

S

etup.AutoInfo.P

on, off

S

etup.AutoInfo.Ch.G

on, off

S

etup.AutoInfo.Ch.GC

on, off

S

etup.AutoInfo.Ch.R

on, off

S

etup.AutoInfo.Ch.S

on, off

S

etup.AutoInfo.Ch.H

on, off

S

etup.AutoInfo.Ch.C

on, off

S

etup.AutoInfo.Ch.B

on, off

S

etup.AutoInfo.Ch.F

on, off

S

etup.AutoInfo.Ch.OM

on, off

S

etup.AutoInfo.Ch.CM

on, off

S

etup.AutoInfo.E

on, off

"on" means, the Sample changer automatically sends a corre-
sponding message when the specified event occurs.
.Status

Enables/disables the preset AutoInfo messages

.P

PowerOn: Simulation of PowerOn (see 4.10.2.43).
Not from mains.

Messages from changer functions:
.Ch.G

Go: Method has been started

.Ch.GC

Go Command: Start command has been received

.Ch.R

Ready: Status 'Ready' reached

.Ch.S

Stop: Status 'Stop' reached

.Ch.H

Hold: Status 'Hold' reached

.Ch.C

Continue: Resuming after Hold

.Ch.B

Begin: Begin of sample sequence

.Ch.F

Final: End of sample sequence