Metrohm 730 Sample Changer Manuel d'utilisation

Page 132

Advertising
background image

4.9 Operation via RS232 Interface

730 Sample Changer, Detailed Description

129

Program-specific Messages:

E28

Wrong object call

E29

Wrong value or no value allowed

E30

Wrong trigger

RS Receive Errors:

E36

Parity error
Exit: <QUIT> and set the same parity for both instruments.

E37

Stop bit error
Exit: <QUIT>and set the same stop bit for both instruments.

E38

Overrun error. At least 1 character could not be read.
Exit: <QUIT>.

E39

Internal receive buffer full (>82 characters).
Exit: <QUIT>.

RS Send Errors:

E40

DSR=OFF. No proper handshake for more than 1s. Exit: <QUIT>.
Is the receiver switched on and ready to receive?

E41

DCD=ON. No proper handshake for more than 1s. Exit: <QUIT>.
Is the receiver switched on and ready to receive?

E42

CTS=OFF. No proper handshake for more than 1s. Exit: <QUIT>.
Is the receiver switched on and ready to receive?

E43

Transmission has been interrupted for at least 3 s with XOFF.
Exit: send XON or <QUIT>.

E44

The RS-interface parameters are no longer identical for both in-
struments. Re-set.

E45

The receive buffer of the Sample Changer contains an incomplete
command (LF missing). Transmission is therefore blocked.
Exit: send LF or <QUIT>.

E50…E59

I/O-Test error

E60…E82

RS232-Test error

Instrument-specific Errors:

E201

Sample changer function error

E202

Dosimat function error