4.10 Remote Control Commands
730 Sample Changer, Detailed Description
151
&Assembly …
4.10.2.50
A
ssembly.Sample
$G
A
ssembly.Sample.Func
=
, +, –
A
ssembly.Sample.Value
1
...999
Defines the (first) sample beaker (rack position) to be processed.
Modification of the sample variable.
.Func
Function
.Value
Value (absolute or relative)
&Assembly.Sample;$G triggers this function.
4.10.2.51
A
ssembly.Move
$G, $S
A
ssembly.Move.Target
1, 2
A
ssembly.Move.Position
sample
, spec.1…8, 0...999
Positioning a beaker at the specified tower.
.Target
Target or tower
.Position
Rack position or identification of beaker
&Assembly.Move;$G triggers this function.
4.10.2.52
A
ssembly.Lift
$G, $S
A
ssembly.Lift.Station
1
, 2,
∗
A
ssembly.Lift.Way
rest
, work, rinse, shift, special, 0...325 mm
Move lift.
.Station
Selection of lift station (
∗
= both lifts)
.Way
absolute lift position
&Assembly.Lift;$G triggers this function.
4.10.2.53
A
ssembly.Stir
$G, $S
A
ssembly.Stir.Adress
1
...4,
∗
A
ssembly.Stir.Value
1
...9999 s, on, off
Stirrer control.
.Address
Selection of stirrer (
∗
= all stirrers)
.Value
State or time interval in sec.
&Assembly.Stir;$G triggers this function.
4.10.2.54
A
ssembly.Pump
$G, $S
A
ssembly.Pump.Adress
1.1
, 1.2, 1.
∗
, 2.1, 2.2, 2.
∗
A
ssembly.Pump.Value
1
...999 s, on, off
Pump control. Only two pumps may be operated concurrently.
.Address
Selection of pumps in format 'tower.pump'
(
∗
= both pumps of a tower)
.Value
State or time interval in sec.
&Assembly.Pump;$G triggers this function.