background image

4.10 Remote Control Commands

730 Sample Changer, Detailed Description

151

&Assembly …

4.10.2.50

A

ssembly.Sample

$G

A

ssembly.Sample.Func

=

, +, –

A

ssembly.Sample.Value

1

...999

Defines the (first) sample beaker (rack position) to be processed.
Modification of the sample variable.
.Func

Function

.Value

Value (absolute or relative)

&Assembly.Sample;$G triggers this function.

4.10.2.51

A

ssembly.Move

$G, $S

A

ssembly.Move.Target

1, 2

A

ssembly.Move.Position

sample

, spec.1…8, 0...999

Positioning a beaker at the specified tower.
.Target

Target or tower

.Position

Rack position or identification of beaker

&Assembly.Move;$G triggers this function.

4.10.2.52

A

ssembly.Lift

$G, $S

A

ssembly.Lift.Station

1

, 2,

A

ssembly.Lift.Way

rest

, work, rinse, shift, special, 0...325 mm

Move lift.
.Station

Selection of lift station (

= both lifts)

.Way

absolute lift position

&Assembly.Lift;$G triggers this function.

4.10.2.53

A

ssembly.Stir

$G, $S

A

ssembly.Stir.Adress

1

...4,

A

ssembly.Stir.Value

1

...9999 s, on, off

Stirrer control.
.Address

Selection of stirrer (

= all stirrers)

.Value

State or time interval in sec.

&Assembly.Stir;$G triggers this function.

4.10.2.54

A

ssembly.Pump

$G, $S

A

ssembly.Pump.Adress

1.1

, 1.2, 1.

, 2.1, 2.2, 2.

A

ssembly.Pump.Value

1

...999 s, on, off

Pump control. Only two pumps may be operated concurrently.
.Address

Selection of pumps in format 'tower.pump'
(

= both pumps of a tower)

.Value

State or time interval in sec.

&Assembly.Pump;$G triggers this function.