5.4 Diagnosis

730 Sample Changer, Appendix

172

5.4.7 RS232-Interface

RS232-Interface

This diagnosis step tests the functionality of the serial interface.

Prepare instrument for diagnosis (see chapter. 5.4.2).

If necessary, press <ê> several times until

diagnosis

>RS232 test

<ENTER>

>RS232 test

RS232 test connector ?

Insert the test plug 3.496.8480 into the RS232 interface without
turning off the instrument.

<ENTER>

The test runs automatically. If no error occurs, the following appears:

>RS232 test

RS232 test ok

Remove test plug and press <ENTER>.

diagnosis

>external bus test

5.4.8 External Bus Interface

This diagnosis step tests the functionality of parts of the external
bus interface.

Prepare instrument for diagnosis (see chapter 5.4.2).

If necessary, press <ê> several times until

diagnosis

>external bus test

<ENTER>

The test searches for all instruments that are connected on the E-bus interface. The sam-
ple changer considers the following to be “instruments”:

Tower 1 (minimum version)
Tower 2 (optional)
Dosimat Interface (optional, maximum 3 pieces)

If no error occurs, an identification for each instrument which is connected on the E-bus
must appear on the display.

You can browse through this list of identifications found by pressing the <ENTER> key. The
following table shows which identifications must appear with which configurations.