4.10 Remote Control Commands

730 Sample Changer, Detailed Description

146

4.10.2.30

I

nfo.ActualInfo.Pump.1.State

etc. up to .4
Current pump state (on/off).
Pump 1 and 2 on tower 1, Pump 3 and 4 on tower 2.

4.10.2.31

I

nfo.ActualInfo.Buret.1.State

read only

I

nfo.ActualInfo.Buret.1.Position

read only

I

nfo.ActualInfo.Buret.1.Cock

read only

I

nfo.ActualInfo.Buret.1.Type

read only

I

nfo.ActualInfo.Buret.1.Volume

read only

etc. until .12
Current data of dosing drive.
State:

Status (ready/busy)

Position:

Piston position in mL

Cock:

Cock position

Type:

Type of dosing unit (685/700)

Volume:

Burette volume

4.10.2.32

I

nfo.ActualInfo.Inputs.Status

read only

Status of the Input lines (Input0…7) of the Remote interface.
$Q sends the signal state as decimal number e.g.
10

00001010 binary

2

1

+ 2

3

Input1 and Input3 active

(active = low, inactive = high)
See also page 119ff.

4.10.2.33

I

nfo.ActualInfo.Outputs.Status

read only

Status of the output lines (Output0…13) of Remote interface. See
4.10.2.32.

4.10.2.34

I

nfo.ActualInfo.Display.L1

read only

I

nfo.ActualInfo.Display.L2

read only

Text of the first and second line of the LCD display.

4.10.2.35

I

nfo.ActualInfo.Counter.Sample

read only

I

nfo.ActualInfo.Counter.Maximum

read only

Currently processed sample number and total number of samples.