5.4 Diagnosis

730 Sample Changer, Appendix

173

Configuration

Identification in the Display

Tower 1

address 0x86 type 3

Tower 2

address 0x87 type 3

Dos 1 - 4

(729 / Address 1)

address 0x91 type 3

address 0x92 type 3

Dos 5 - 8

(729 / Address 2)

address 0xA1 type 3

address 0xA2 type 3

Dos 9 - 12 (729 / Address 3)

address 0xB1 type 3

address 0xB2 type 3

>external bus test

address 0x86 type 3

Press <ENTER> several times and compare display with the
list above until

diagnosis

>beeper test

5

5.4

.4.9

.9 Beeper

Prepare instrument for diagnosis (see Chap. 5.4.2).

If necessary, press <ê> several times until

diagnosis

>beeper test

<ENTER>

The beeper is turned on and off in an endless loop.

>beeper test

The test can be exited by pressing the <QUIT> or <STOP>
keys.

diagnosis

>rack code test

5.4.10 Rack Code Recognition

Rack Code Recognition

This diagnosis step checks the functionality of the sensors with re-
gard to their ability to automatically recognize the attached rack.

Note the code (magnets) on the underside of all the racks being
used. The following diagram shows the magnet fixture on the
underside of a rack.