4.6 Dosimats and Dosinos

730 Sample Changer, Detailed Description

111

The maximum dosing and filling rates that can be entered in the
configuration menu under '>dosing units' for every port of a dosing
unit depend on the burette size:

Volume of the

dosing unit

Max. dosing rate

Resolution

2 mL

7 mL/min

0.2

µ

L

5 mL

17 mL/min

0.5

µ

L

10 mL

33 mL/min

1.0

µ

L

20 mL

67 mL/min

2.0

µ

L

50 mL

160 mL/min

5.0

µ

L

The following commands can be executed with Dosinos. The as-
sociated inlets and outlets (ports) can be defined in the Parameter
Menu under '>dosing unit def.' as standard arrangement for a par-
ticular method or with a DEF command in a process sequence or
manual operation.

Dosing

Dosing

DOS: XX : yyy.yy ml

Dosing a certain volume.

The volume indicated is ejected at the dosing port. The burette is
not refilled after every operation. The dosing port can be redefined
as desired:

>dosing unit def. XX

dosing port Y

or

<DEF>

DRIVE.PORT XX.Y : dos.

Filling

Filling

DOS: XX : fill ml

Filling the Dosino burette.

The burette is completely filled. It is aspirated via the filling port.
This can be redefined as desired:

>dosing unit def. XX

filling port Y

or

<DEF>

DRIVE.PORT XX.Y : fill