5 initialize data memory – Metrohm 730 Sample Changer Manuel d'utilisation

Page 178

Advertising
background image

5.5 Initialize data memory

730 Sample Changer, Appendix

175

5.5 Initialize data memory

This diagnostic step can be used to write default values to the in-
strument parameters using the keypad and thus switch the instru-
ment to the original condition. This measure is important with the
following two points:

The setting of certain instrument parameters such as the locking of

keys is possible only via RS232, i.e. with the aid of a PC. If such in-

strument parameters are set and no PC is available to cancel the

settings, full use can not be made of the instrument.

In rare cases, it is possible that major interference signals such as

line spikes and lightning can have an adverse influence on the

contents of the data memory. If the contents of the data memory

are undefined, this may lead to a system crash.

The 730 Sample Changer offers various possibilities to initialize
the data memory. The entire data memory (

all

) or only parts of it

(param, config, setup, assembly

) can be written to with default

values.

Although the instrument number will be retained, the initialization

should be performed only if necessary as the stored user data

(etc.) are deleted.

Prepare instrument for diagnostic test (see chapter 5.4.2).

If need be, press <9> key repeatedly until

diagnosis

>RAM initialization

Press <ENTER> key to open the following diagnostic menu: