Rxd txd, External device, Sample changer – Metrohm 730 Sample Changer Manuel d'utilisation

Page 159: External device sample changer, Time rxd txd, Sample changer external device, Sample changer as receiver, Sample changer as sender, As soon as an l

Advertising
background image

4.11 Properties of the RS232 Interface

730 Sample Changer, Detailed Description

156

Sample Changer as Receiver:

Time

RxD

TxD

Data output

Data input

XON

XOFF

60 characters

max. 22

characters

XOFF

Time of
4 characters

String

processing

400 µs

external device

LF

Sample Changer

Sample Changer as Sender:

Time

RxD

TxD

Data output

Data input

XON

XOFF

max.
4 characters

Data output

disabled

L

F

Data output

enabled

Sample changer

external device

Software-Handshake, SWline

Handshake input ports on the changer (CTS, DSR, DCD) are not
checked.
Handshake output ports (DTR, RTS) are set by the changer.
The changer has an input buffer which can accept up to 80 char-
acters + C

R

L

F

. As soon as an L

F

is recognized, the changer

sends an XOFF. After this, it can receive and save a maximum of
6 characters. The character string previously sent is now proc-
essed by the changer. Afterwards the changer sends XON and is
again ready for receiving.

Sample Changer as Receiver:

XOFF

XON

Time

RxD

TxD

Data output

Data input

Input

disabled

LF

LF

400 µs

String
processing

external device

Sample Changer

Advertising