background image

4.10 Remote Control Commands

730 Sample Changer, Detailed Description

144

4.10.2.21

C

onfig.WetPart.WetPartNo

1

...12

Identification of dosing unit.

4.10.2.22

C

onfig.WetPart.1.MaxRate

0.01...160 ml/min

C

onfig.WetPart.1.Length

0...1000...30000 mm

C

onfig.WetPart.1.Diameter

0.1...2...20 mm

until .4.Diameter
(Default value .2.Length : 250 mm)
Configuration of the dosing units. Depending on the selection of
the dosing unit (see WetPartNo 4.10.2.21), the corresponding data
set is overlayed. These settings are only relevant for Dosinos. For
each port of the Dosino individual settings are possible.
MaxRate:

Max. possible dosing and filling speed allowed

Length:

Tubing length on selected port

Diameter:

Inner tubing diameter on selected port

4.10.2.23

C

onfig.RSset

$G

$G effects all the RS settings. Modifications are only possible in
inactive instrument state. After setting the interface parameters
wait for 2 sec. to allow the components to equilibrate.

4.10.2.24

C

onfig.RSset.Baud

300, 600, 1200, 2400, 4800, 9600

C

onfig.RSset.DataBit

7, 8

C

onfig.RSset.StopBit

1

, 2

C

onfig.RSset.Parity

even, odd, none

C

onfig.RSset.Handsh

HWs

, HWfull, SWchar, SWline, none

Settings for data transmission via RS interface, baud rate, data bit,
stop bit, parity and type of handshake, see also page 155ff.

4.10.2.25

C

onfig.RSset.CharSet

IBM

, HP, Epson, Seiko, Citizen

Setting the character set and print mode. For data communication
with computers select 'IBM' (IBM code page 437).