3.5 Methods and Sequences

730 Sample Changer, Introduction

51

Procedure for creating a method:

Enter a command or select an existing command line

Press the <LEARN / HOLD> key

Function is started, "LEARN" LED lights up

Press the <LEARN / HOLD> key

Function is stopped, "LEARN" LED blinks

With the <ENTER> key, accept the value (or re-start the
LEARN function)

"LEARN" LED goes out, edit next command line

The LEARN function can be used with the following commands:

Command "Teachable" Parameter

Mode of function

LIFT

Lift position in mm

absolute

PUMP

Pump time in sec

additive

STIR

Stirring time in sec

additive

WAIT

Waiting time in sec

additive

DOS

Dosing volume in mL

additive

SCN Rm Status of the 8 remote lines

"live" value

SCN RS

Character sequence received

"live" value

TRACE function

The "TRACE" function is a valuable aid for operating through an
entire sequence or method (or parts thereof) for test purposes.
Every command line in a sequence can be executed directly by
pressing the <START> key. Upon completion of the action the
next command line is displayed.
Tracing can be executed immediately after entry of a sequence
line or at any time after opening the parameter menu and selecting
a sequence.

3.5.3 Process Control

With <START> a method is started from the normal state. If there
is no manual intervention or unexpected errors, the sample series
is correctly processed and closed with the final sequence. The
sample sequence is executed repeatedly according to the entry
under 'number of samples', beginning with the sample beaker that
is defined as 'SAMPLE'.

If the sample series is interrupted with <STOP>, the sample
changer returns immediately to the normal state. Samples that
have not been processed are ignored and the end sequence is not
executed. If settings for such instances have been activated under
'manual stop options'

, the corresponding actions or commands