5.4 Diagnosis

730 Sample Changer, Appendix

169

5.4.3 Working Memory (RAM)

This diagnosis step completes a non-destructive test over the en-
tire range of the RAM contents (working memory).

Prepare instrument for diagnosis (see chapter 5.4.2).

If necessary, press <ê> several times until

diagnosis

>RAM test

<ENTER>

If no errors are found, this appears on the display:

>RAM test

RAM test ok

<ENTER>

diagnosis

>display test

5.4.4 Display

With this diagnosis step, the functionality of the LED’s and the dis-
play is tested.

Prepare instrument for diagnosis (see chapter 5.4.2).

If necessary press <ê> several times until

diagnosis

>display test

<ENTER>

After pressing the <ENTER> key, the program automatically makes a test run to visually
check the LED’s and the display.
⇒ The LEDs for TOWER 1, TOWER 2 and LEARN blink one after the other for a short

time.

⇒ The background lighting of the display is switched off for a short time and then turned

on again.

⇒ The start test pattern appears (every pixel active).
⇒ Both lines of the display are turned off.
⇒ Both lines of the display are occupied one after the other with the characters “#”, “H

and finally “I”.

⇒ Both lines are filled from right to left with the endless running line

“0123456789ABCDEFGHIJKLMNOPQRSTUVWXYZ.”

The test run can be stopped and started again by pressing the
<5> key.

The test can be exited with the <QUIT> or <STOP> keys.