5.5 Initialize data memory

730 Sample Changer, Appendix

176

>RAM initialization
select: param

ü

>RAM initialization
select: config

ü

>RAM initialization
select: setup

ü

>RAM initialization
select: assembly

ü

>RAM initialization
select: all

ь

ь

ь

ь

ь

Set method parameters with default values.

Ã

Set config parameters with default values.

Ã

Set setup parameters with default values.

Ã

Set assembly parameters with default values.

Ã

Deletes all user-defined methods.

Ã

Pressing the <SELECT> key selects the submenus in turn. The
individual initialization alternatives are accessed using the
<ENTER> key, exit is with the <QUIT> key.

The table shows which parts of the data memory are affected by
the corresponding initialization alternatives. When the instrument
reacts with a system crash after being switched on (undefined dis-
play, no reactions to keystroke, etc.) we recommend to carry out
the initialization with the submenu "

all

".

If need be, press the <SELECT> key repeatedly until:

>RAM initialization

select: all

<ENTER>.

diagnosis

>RAM test

<QUIT>.

The instrument quits the diagnostic menu and runs a power on reset.