4.10 Remote Control Commands

730 Sample Changer, Detailed Description

154

&Diagnosis …

4.10.2.61

D

iagnose.Init

$G

D

iagnose.Init.Select

param, config,

setup, assembly, all

RAM initialization. Sets all default values for the selected sub-
branch. See also 4.10.2.44.
&Diagnose.Init;$G triggers the initialization.

4.10.2.62

D

iagnose.RamTest

$G

D

iagnose.LcdTest

$G, $S, $H

D

iagnose.ContrastTest

$G, $S

D

iagnose.KeyTest

$G, $S

D

iagnose.IoTest

$G, $S

D

iagnose.RsTest

$G, $S

D

iagnose.EbusTest

$G, $S

D

iagnose.BeeperTest

$G, $S

D

iagnose.RackcodeTest

$G, $S

D

iagnose.FunctionTest

Diagnostic functions. These functions can be triggered with $G.
They may be stopped with $S.
.RamTest

Working memory test

.LcdTest

Display test

.ContrastTest Display contrast test
.KeyTest

Displays key code and function of each key

.IoTest

Remote interface test

.RsTest

(This function is not applicable via RS232 inter-
face)

.EbusTest

External Bus interface test

.BeeperTest

Test of the acoustic signal (only with
Config.Beeper"on", see 4.10.2.14)

.RackcodeTest Test of rack code recognition
.FunctionTest Metrohm internal test

4.10.2.63

D

iagnose.SimulateKey

0

, 1…6, 8…31

Key simulation (Key codes see 4.10.2.37)

4.10.2.64

D

iagnose.InstrNo

Instrument number. This entry is not accessible via RS232 inter-
face.

4.10.2.65

D

iagnose.PowerOn

$G

Simulation of 'Power on'.

Advertising