3.1 Tutorial

730 Sample Changer, Introduction

28

Insert some sample beakers into the sample rack. One should
be put in the highest possible rack position as rinse or condi-
tioning beaker. The rest of the beakers are placed in ascending
order, beginning with position 1. Using the keys <ç> and <è>
the rack can be turned for this purpose.

Basic Configuration

Move the lift downwards carefully with the <ê> key so that the
electrode being used almost touches the bottom of the titration
beaker or the titration head almost touches the edge of the
beaker.

CONFIG

ENTER

TOWER

SEL ECT

The dialog language can be set in the configuration menu.
Press <CONFIG>

Display:

configuration

>auxiliaries

and then <ENTER>.

Display:

>auxiliaries

dialog: english

This menu item has a colon, indicating that here the parameters
can be selected from a list. Press <SELECT> several times in
order to view the various selections and get used to this type of
dialog.

Display:

>auxiliaries

dialog: deutsch

dialog: francais

With <ENTER> you can accept the suggestion 'dialog: english'.

Display:

>auxiliaries

display contrast: 3