2 description of the remote control commands – Metrohm 730 Sample Changer Manuel d'utilisation

Page 142

Advertising
background image

4.10 Remote Control Commands

730 Sample Changer, Detailed Description

139

4.10.2 Description of the Remote Control Commands

&Mode …

4.10.2.1

M

ode

$G, $S, $H, $C

Start ($G) or stop ($S) the current method. Interrupt with $H
(hold), resume with $C (continue).

4.10.2.2

M

ode.Method

read only

Name of current method in the working memory. $Q sends
8 ASCII characters.

∗∗∗∗∗∗∗∗

stands for a blank method.

4.10.2.3

M

ode.SmplNo

1...999,

, rack

Number of samples. This entry determines the number of runs for
the sample sequence.

= infinite number of samples. Processing has to be stopped

with &M;$S or <STOP>.

rack = (Number of rack positions – number of special beakers

defined).

4.10.2.4

M

ode.StartSeq.1.Cmd

NOP

, MOVE, LIFT, STIR,

etc. up to .99

DEF, PUMP, DOS, SCAN,

SAMPLE, CTRL, WAIT, ENDSEQ

Defines the command of the indexed command line in a start se-
quence. The introduction of a command appends the according
sub-branch from &Assembly (see page 136) to the index node.
A NOP-entry cuts the appended sub-branch from the index node.
Each entry at the end of a sequence appends a new node
&Mode.StartSeq.

.Cmd("NOP").

4.10.2.5

M

ode.StartSeq.1.

.Move…, .Lift…, .Stir…, .Pump…,

etc. up to .99

.Dos…, .Scan…, .Ctrl…, .Def…,

.Sample…, .Wait…, .End

Indexed start sequence; its commands will be executed line by line
in processing. At the index node &Mode.StartSeq.1.Cmd (see
4.10.2.4) the according sub-branch from &Assembly… will be ap-
pended depending on the selected command (see 4.10.2.4).
Example:
&Mode.StartSeq.1.Cmd("MOVE")

&Mode.StartSeq.1.Move.Target("1")

&Mode.StartSeq.1.Move.Position("sample")

Advertising