4.3 Menu Organization

730 Sample Changer, Detailed Description

88

parameters

>start sequence

Editor for the start sequence of a sample series

Open the submenu with <ENTER>

The processing sequence entered here is executed once at the
start

of a sample series. This can be useful for rinsing or condi-

tioning the electrode, for example.

parameters

>sample sequence

Editor for processing sequence for each sample

Open the submenu with <ENTER>

This processing sequence is executed during the processing of
every sample

of a series.

parameters

>final sequence

Editor for the end sequence of a sample series

Open the submenu with <ENTER>

This processing sequence is executed once at the end of a sam-
ple series. This could be the positioning of a rinsing or conditioning
beaker, for example.

In principle the same rules for entry are valid here as for manual
operation, i.e. after selecting a command and entering the neces-
sary data, the entry is accepted with <ENTER>. Afterwards the
next command line is accessed where a new command can be
entered.

The "LEARN" mode is available for the easy entry of parameters.
With certain commands "live" values can be taken over by manual
execution of a single command. More information can be found on
page 50.

Furthermore the "TRACE" function can be used to execute every
command line step by step. See page 50.

Navigation in a sequence is accomplished as in the other menus.
In addition the <INSERT> and <DELETE> keys can be used.

<INSERT> adds a new command line above the current line in a
sequence. It is automatically occupied by the "NOP" command
that has no function. The following lines are shifted one line
downwards.

<DELETE> deletes the current line in a sequence. The lines fol-
lowing are shifted one line upwards.