5.1 Error Messages

730 Sample Changer, Appendix

161

5 Appendix

Appendix

5.1 Error Messages

When an error occurs the execution of the active command is in-
terrupted and an error message is displayed (display blinks). This
must be acknowledged with the <QUIT> key.
If an error occurs during processing of a sample series, the
changer will then be switched into the 'HOLD' state. After the
cause of the error has been rectified, the sample series can be
continued with the next command in the active sequence by
pressing the <START> key. If the error cannot be eliminated, the
method running can also be halted with <STOP>.

The list of possible error messages and their causes:

* battery low

The battery for the permanent storage of user meth-
ods must be replaced.

* changer low power

The power supply cannot deliver enough power for
the simultaneous operation of all components cur-
rently in use (stirrers, pumps, lifts).

* changer not ready

The changer cannot execute the command chosen
because it is busy carrying out another action.

* changer overload

Load or resistance too large to carry out the chosen
action.

* Dos## command not exec

An error has occurred with the indicated dosing unit.

* Dos## ext. bus failure

An unexpected error has occurred with an instrument
(Dosimat, Dosino) that is connected on the external
bus.

* Dos## no exchange unit

There is no exchange unit mounted on the dosing
unit indicated.

* dos.unit ## missing

The dosing unit chosen is not connected.