2.5 Integration

730 Sample Changer, Installation

16

Remote control commands:

CTL:Rm : START device1

starts measuring instrument

CTL:Rm : *************1

"

CTL:Rm : METER mode pH

change mode to pH measurement

CTL:Rm : *********0001*

"

CTL:Rm : METER mode T

change mode to temp. measurement

CTL:Rm : *********0010*

"

CTL:Rm : METER mode U

change mode to mV measurement

CTL:Rm : *********0011*

"

Scan commands:

SCN:Rm : End1

awaits end of measurement

SCN:Rm : ****1***

"

730 Sample Changer — pH- / Ion- / Conductometer
(692/712/713)

692
712
713

730

Remote control commands:

CTL:Rm : START device1

starts measuring instrument

CTL:Rm : *************1

"

CTL:Rm : METER mode pH

change mode to pH measurement

CTL:Rm : *********0001*

"

(not appl. with 712)

CTL:Rm : METER mode T

change mode temp. measurement

CTL:Rm : *********0010*

"

(not appl. with 712)

CTL:Rm : METER mode U

change mode to mV measurement

CTL:Rm : *********0011*

"

(not appl. with 712)

CTL:Rm : METER mode I

change mode to Ipol (mV meas.)

CTL:Rm : *********0100*

"

(not appl. with 712)

CTL:Rm : METER mode C

change mode to conc measurement

CTL:Rm : *********1000*

"

(only appl. with 692)

CTL:Rm : METER cal pH

change mode to pH calibration

CTL:Rm : *********0101*

"

(not appl. with 712)

CTL:Rm : METER cal C

change mode to conc calibration

CTL:Rm : *********1001*

"

(only appl. with 692)

CTL:Rm : METER enter

simulates <ENTER> key

CTL:Rm : *********1111*

"

(not appl. with 712)

Scan commands:

SCN:Rm : end1

awaits end of determination

SCN:Rm : ****1***

"

Cable 6.2141.020