4.11 Properties of the RS232 Interface

730 Sample Changer, Detailed Description

159

4.11.3 Pin Assignment

RS232C Interface

Transmitted Data (TxD).

If no data are transmitted, the line is held in the “ON”

condition. Data will only be sent when CTS and DSR are

in the ”ON” condition and DCD is in the “OFF” condition.

Received Data (RxD)

Data will only be received when DCD is “ON”.

Request to Send (RTS)

ON condition: Sample Changer is ready to send data.

Clear to Send (CTS)

ON condition: Remote station is ready to receive data.

Data Set Ready (DSR)

ON condition: The transmission line is connected.

Signal ground (GND)

Data Carrier Detect (DCD)

ON condition: The level of the received signal is within the

tolerance range (remote station is ready to send data).

Data Terminal Ready (DTR)

ON condition: Instrument is ready to receive data.

Fehler! Keine gültige Verknüpfung.

Protective earthing
Direct connection from cable plug to the protective ground of the instrument.

Polarity allocation of the signals
-

Data lines (TxD, RxD)

voltage negative (<–3 V): signal state "ON"
voltage positive (>+3 V): signal state "ZERO"

-

control or message lines (CTS, DSR, DCD, RTS, DTR)

voltage negative (<–3 V): OFF state
voltage positive (>+3 V): ON state

In the transitional range from +3 V to –3 V the signal state is undefined.

Driver 14C88 according to EIA RS232C specification
Receiver 14C89

" "

Contact arrangement at plug (female) for RS232C socket (male)