4.8 Remote Interface

730 Sample Changer, Detailed Description

123

SCN Command

Parameter

Bit Pattern

Function

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
ready1

*******1

queries "ready" state of device1 (Titrinos)

ready2

**1*****

queries "ready" state of device2 "

ready*

**1****1

queries “ready" state of device1 and 2 "

end1

****1***

expects the end pulse of device1 (for ex. EOD)

end2

*1******

expects the end pulse of device2 “

endmeter

***11***

expects the end pulse from 692 Ion Meter or
713 pH Meter (during the waiting interval, the
stirrer 1 is turned on)

With the 'ready

' parameter the readiness of two concurrently operating instruments

may be queried. The 'ready' line of both instruments must be automatically set (static)
at the end of a determination. Two instruments that emit short pulses at the end of e.g.
measurement, can not be controlled at the same time.