Les performances relatives aux arythmies sont rapportées conformément à l’article, Kerber RE,
Becker LB, Bourland JD, Cummins RO, Hallstrom AP, Michos MB, Nichol G, Ornato JP,
Thies WH, White RD, Zuckerman BD. « Automated External Defibrillators for Public
Access Defibrillation : Recommendations for Specifying and Reporting Arrhythmia
Analysis Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety, »
Circ J Am Heart Assoc. 1997;95:1677-1682.

Références

Young KD, Lewis RJ. « What is confidence ? Part 2 : Detailed definition and determination
of confidence intervals ». Ann Emerg Med. September 1997 ; 30:311-318.

Beyer WH. Percentage Points, F-Distribution Table. CRC Standard Mathematical Tables.
28

th

ed. Boca Raton, Fla : CRC Press ; 1981:573.

Tableau A-5.

Résultats relatifs aux performances cliniques (patients enfants)

Rythmes

Taille

échantillon

(enregistrements

de 9 secondes)

Objectifs

de

performance

Performance

observée

Limite

inférieure

de l’intervalle

de confiance

à 90 %

unilatéral

Traitable par choc
(49 patients)

Sensibilité

FV brute

42

> 90 %

100,0 %

93,1 %

TV rapide

79

> 75 %

100,0 %

96,3 %

Non traitable par choc
(155 patients)

Spécificité

RSN

208

> 99 %

100,0 %

98,6 %

FA, BS, TSV

a

,

bloc cardiaque,
idioventriculaire, ESV

a. 161 des 348 rythmes anormaux enregistrés étaient TSV (72 patients). Les fréquences cardiaques TSV étaient

comprises entre 152 et 302 battements par minute.

348

> 95 %

99,4 %

98,2 %

Asystolie

29

> 95 %

100 %

90,2 %

Intermédiaire
(16 patients)

FV fine

0

Uniquement

rapportées

Autres TV

44

Uniquement

rapportées

84,1 %

72,2 %