Chapter 6: Maintaining the Protector Echo

Product Service 1-800-522-7658

91

Original instructions

(You will need this information only if you have gGuard installed.)

If DHCP is

yes

then you have a dynamic IP if

no

then you have a static IP which

you must reenter completely.

Elapsed Time
Filter X D (this is the # of days since install)

Fans
Set point Dflt. XXXX rpm
Set point XXXX rpm

Semicon S. trig value (example 3500)
Trig. X
Delay X

Sash
Which type (example. normal polarity)
Trig X %


Now press

“X”

to return to the previous menu.”

Step 4 – Turn off the hood

Press the fan icon to turn off the controller waiting the 30 seconds to completely
shut down. Also if the light is on, turn it off as well.

Step 5 – Remove failed fan box

Take off the secondary filter that is directly above the fan you are replacing. Once
you have it removed unplug the following:

The left fan power supply wire

Air sampling tube

The right fan power supply wire to the next fan (if applicable)


***Again make note of the

micro

switch settings of that fan.***

CAUTION: Do not contact fan wheel while still in motion.

PRUDENCE: Ne pas contacter roue du ventilateur tout en en
mouvement.

!