Chapter 6: Maintaining the Protector Echo

Product Service 1-800-522-7658

77

Original instructions


Take them down one by one. For each, lift up on the front and then turn your hand
left then right to break the back edge free from the filter. Once you have the back
of the fan free, you can pull it towards you until you are able to grab it with your
other hand.

Once you have it with both hands either hand it to someone else standing nearby
or be very careful and step down the ladder.

*** Be very careful not to chip the plastic sealing edge of the fan boxes.***

Now as you face the hood looking left to right, you should see your primary filters
marked “used” and your lights below.

Now go back to the control module press

“OK”

to illuminate the screen. Then

press

“OK”

again to go to the next step.


Step 9 – Remove your old primary filters that you marked “used”

You should now see:

[Partial Replacement]

Step #3

Remove Old Filters


Take down your old primary filters and put in a stack near the hood (as you will
need to scan them out later).

Take them down one by one. For each filter, lift up on the front and then turn your
hand left then right to break the back edge free from the filter. Once you have the
back of the filter free, pull it towards you until you are able to grab it with your
other hand.

Once you have it with both hands either hand it to someone else standing nearby
or be very careful and step down the ladder.

At this time, it is strongly recommended that you inspect your lights and pre-
filters since this is a convenient time to replace your pre-filters and swap out any
bad or suspect bulbs.

Advertising