Product Service 1-800-522-7658

65

Original instructions

Quarterly (or more often as required)

Determine the actual airflow face velocity through the sash opening of the

filtered hood where the average reading should be at the specified
velocity. (Use calibrated thermal anemometer or other approved
apparatus).

Replace carbon filters when chemical breakthrough is indicated by the

primary broad organic sensor, target acid gas sensor or time. See “Partial
Filter Replacement” section of this manual in Chapter 6.

All weekly activities.

Annually

Replace lights, if necessary.

Replace HEPA filters if the face velocity drops below the recommended

speed as the airflow alarm alerts you.

Have the enclosure validated by a qualified certification technician. See

Certification and Recertification in Chapter 6.

All quarterly activities.

Initial Certification


The filtered hood has been designed for an inflow velocity of 60-100 fpm. The
filtered hood should be certified for the proper inflow velocity required by your
Safety Officer.

Re-Certification


Under normal operating conditions, the filtered hood should be recertified at least
annually if serviced. The certifier should perform the following tests.

Airflow Velocity Test


In addition, the following test should also be performed at the user’s discretion:

Smoke Test to determine proper airflow patterns

User Alarm Code List


Listed next is an overview of control panel functions and alarm codes with their
meaning, causes, and solutions.