Chapter 3: Getting Started

Product Service 1-800-522-7658

39

Original instructions

You should now see:

>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases
Audible alarm Mgmt.


Arrow down to

“Fan set point”

press

“Ok”

to select it.


You should now see:

>Default (2850)
Minimal (1500)

Use the up or down arrows to put in

“X”

from your notes in step 2 or refer to

GFH Settings sheet.

“Confirm new value”

press

“Ok”

to confirm.


Arrow down to

“Minimal”

press

“Ok”

to select it.

Use the up or down arrows to put in

“X”

from your notes in step 2 or refer to

GFH Settings sheet.

“Confirm new value”

press

“Ok”

to confirm.


Place your Card back into the card reader.

Press

“X”

to return to the previous menu.


Step 11 – Set alarm delays and choose the sash t

ype.


You should see:

>Hood integration
Fan set point
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash sensor trigger
Software releases
Audible alarm Mgmt.


Arrow down to

“Semicon S. alarm Delay”

press

“Ok”

to select it.

You will see:

05 Minutes
1 Min
60 Max

Make the

0

a

1

for

15 Minutes

Press

“Ok”

then

“Confirm new value”

press

“Ok”

to confirm.