Chapter 3: Getting Started

Product Service 1-800-522-7658

43

Original instructions


Example of static IP (needed for gGuard)

Ethernet network

DHCP: No
IP: 192.168.0.20
MSK: 255.255.255.0
GW: 192.168.0.1

Elapsed times

Filter XD

***Note** It will be 0 as the clock starts over at the time of

Reconfiguration (the memory is wiped clean). It displays the current days on the
filters, so for example tomorrow will show 1 then the next day a 2. However, the
CPU is still counting down the days from the number that you entered from step 5.

Example of fan speeds

Fans

Dflt 2850 rpm
Set point 1500 rpm Variable Volume, 2850 rpm Constant Volume

These values should be the ones that you entered earlier from your notes in step 2.

You should then see an rpm value (make sure hood is on and press

“Ok”

to

refresh) for all the fans that your hood has based on the hood size (either a M2,
M3, M4 or M5).

Example, M4 will have four fans. Your numbers will be different based on your
parameters.

Fan 1; 1533
Fan 2: 1522
Fan 3: 1511
Fan 4: 1630
Fan 5: -------


Example

Semicon Sensor

Trig. M 3500 MV
Delay 15mn


Electro Sensor

Delay 15 mn


Sash
Type Inverted polarity
Trig. 65%


Now check the audible alarm for the sash by raising the sash so it sounds. The
height you hear it should be the upper end of the safe working height above the
sash stop.