Labconco Protecto Ech Filtered Fume Hoods 183 Series 8' Manuel d'utilisation

Page 79

Advertising
background image

Chapter 6: Maintaining the Protector Echo

Product Service 1-800-522-7658

73

Original instructions

Step 3 – Mark your Primary and Secondary Filters

Mark your primary filters (your primary filters are the lowest set of filters from
left to right facing the hood and they are in between the lights and the fans). Write
the word “used” on the primary filter with a marker.

Mark your secondary filters (your secondary filters are the top most set of filters
from left to right facing the hood). Write “primary” on each secondary filter.

Step 4 – Turn off the hood

Press the fan icon once on the control module to turn it off. It will count down 30
seconds and then power down.

***Note*** Leaving the fan on will not allow you to see the filter menu to do a
partial replacement. It will be dashed out.

Step 5 – Select “filters” in the control module

Use your Administrator card and place it into the card reader.

You will now see:

>Hood Status menu
Administration menu


Move down to the

Administrator

menu and press

OK

to select it.


You will now see:

Language
Date and Time
Filters
Alarms and Sensors
Users and Management
Network Parameters
Erase History
Ventilation


Arrow down to

Filters

and press

OK

to select it


***Note*** If at any time during this procedure the screen goes black press

OK

to illuminate it.***

Advertising