Chapter 3: Getting Started

Product Service 1-800-522-7658

32

Original instructions

Step 3 – Remove front panel.

Use your ladder and remove the upper front panel so you can access the RFID
filter tags later in this procedure.


Step 4 – Delete current configuration.

You should now see on the screen:

>Hood Status
Administration menu
Maintenance menu


Arrow down to

“Maintenance Menu”

then press

“Ok”

to select it.


You should now see:

>Hood Integration
Semicon S. Alarm Delay
Electro S. Alarm Delay
Sash Sensor Trig.
Software Releases
Audible Alarm Mgmt.


At

“Hood Integration”

press

“Ok.”


Then arrow down to

“Delete Factory Cfg”

and press

“Ok.”

You should now see

“0000000000.”

Press

“Ok”

until the cursor is on the fourth zero

from the right.

“0000000000”


Now arrow up once to make a

“1”

press

“Ok.”

Arrow up twice to make a

“2”

press

“Ok.”

Arrow up three times to make a

“3”

press

“Ok.”

Arrow up four times to make a

“4.”


You will now have

“0000001234.”


***Note if you make a mistake you can press

“X”

to move back to the left and

change the entry.***

If it is correct then press

“OK.”

You should now see

“Hood will reboot please be patient.”