Appendix E: gGuard
Product Service 1-800-522-7658
133
Original instructions
gGuard Recommendations
To ensure communication, Erlab recommends the following:
gGuard is installed on a PC provided by the user that is on the same network
segment or subnet as the hoods themselves.
If the user PC is not on the same network segment or subnet then another PC
should be installed by the user on the same subnet as the hoods in the lab.
Then 3rd party remote desktop software provided by user should be used to
view that PC with gGuard installed from others on-site that want to view
gGuard on the network.
The PC that has gGuard is configured to never sleep, standby or hibernate
under the power options. This will ensure communication to receive every
email alert from software.
If two or more people would like to receive email alerts then a group email
address should be created by the IT on-site for those who want to get email
alerts from the gGuard software.
If there is an internal Firewall, IT on-site allows FTP transmission through the
firewall, specifically ports 21 and 23.
Static IP’s should be assigned to each hood by IT on-site.