Chapter 6: Maintaining the Protector Echo

Product Service 1-800-522-7658

100

Original instructions

Step 11 – Snake the wires through the hood

Have your helper bring the control module back over to the hood holding the back
of the controller up. Take this time to put the remaining wires up through the hood
as needed.

You should have:

Light supply wire

Fan supply wire

Air tube

Temp sensor

Sash


Step 12 – Connect the electrical cord set to the control Module

If your controller is not hardwired into the electrical system and plugged directly
into an outlet on the top of the hood you may now move on to the next step.

Otherwise if it is hardwired and you must continue to follow the rest of this step.

You should now have the power cord connected to the control module by
reconnecting the quick electrical connector.

Step 13 – Attach the control module to the mounting plate

Take this time to first plug in your ethernet line into the back of the controller (if
applicable). Next have your helper hand it to you. Line up the bottom holes of the
control module to the small arms on the bottom of the front mounting plate.

Once you have it in their slots on the bracket, push the top back to line up the two
screw holes.

***It is very important to have it lined up as the screws can cross thread very
easily.****

This time, place the right screw in first followed by the left. Check that it is
securely fastened on. Once you are sure that they are secure, proceed to the next
step.