Chapter 3: Getting Started

Product Service 1-800-522-7658

31

Original instructions

So that the event can be logged, provide Labconco Product Service with the
serial # of the control module along with the component that failed.


***Note at any time if the LCD screen loses illumination press

“Ok”

and it will light

back up.***

If the

“Hardware Mismatch”

menu is not seen you may find it by opening the

“Hood Status menu.”


Arrow down to

“Hood Status”

and press

“Ok”

to open it.


Step 2 – Write down parameters under “Hood Status.”

In case it is still not shown, press

“Ok”

to open the

“Hood Status”

menu again.


Then arrow up or down to record the following on paper:

DHCP: X

IP: XXX.XXX.XXX.XXX
MSK: XXX.XXX.XXX.XXX
GW: XXX.XXX.XXX.XXX


If gGuard is not installed, skip this step.

If DHCP is

yes

then you have a dynamic IP if

no

then you have a static IP which you

must reenter completely.

Elapsed Time
Filter X D (this is the # of days since install)

Fans
Set point Dflt. XXXX rpm
Set point XXXX rpm

Semicon S. trig value (example 3500)
Trig. X
Delay X

Sash
Which type (example. normal polarity)
Trig X %


Now press

“X”

to return to the previous menu.”