Labconco Protecto Ech Filtered Fume Hoods 183 Series 8' Manuel d'utilisation

Page 82

Advertising
background image

Chapter 6: Maintaining the Protector Echo

Product Service 1-800-522-7658

76

Original instructions

(M3)

Fan #3

Move your ladder to the right so you will be in front of fan #3. Disconnect the air
tube that is on fan #3 and place the air tube on the control module.

Fan #3 should now be free. Proceed to the next step.

(M4, M5)

Fan #3


Move your ladder to the right so you will be in front of fan #3. Leave the wire to
the right that extends to fan #4 connected to fan #3. Unplug the other end from
fan #4.

Disconnect the air tube from the “T” that went to fan #2 and place the tube on top
of the control module. Now pull the air tube off fan #3.

Fan #3 should now be free.

(M4)

Fan #4

Move your ladder to the right so you will be in front of fan #4. Disconnect the air
tube that is on fan #4 and place the air tube on the control module.

Fan #4 should now be free. Proceed to the next step

(M5)

Fan #4


Move your ladder to the right so you will be in front of fan #4. Leave the wire to
the right that extends to fan #5 connected to fan #4. Unplug the other end from
fan #5.

Disconnect the air tube from the “T” that went to fan #3 and place the tube on top
of the control module. Now pull the air tube off fan #4.

Fan #4 should now be free.

(M5)

Fan #5


Move your ladder to the right so you will be in front of fan #5. Disconnect the air
tube that is on fan #5 and place the air tube on the control module.

Fan #5 should now be free. Proceed to the next step.

Step 8 – Remove the fans

*** Be very careful not to chip the plastic sealing edge of the fan boxes.***

Ideally, store them so that the back edge down is on the floor with the circuit
board facing up. *** Place them out of the way.***

Advertising