Chapter 3: Getting Started

Product Service 1-800-522-7658

36

Original instructions

Alarms and Sensors
Users and Management
Network Parameters
Erase History
Ventilation

Arrow to

“Alarms and Sensors”

and press

“Ok”

to select it.

Step 7 – Set semicon trigger value.


You should see:

>Maintenance Alarm
Deactivation Rights
Semicon S. Trig. Value
Sash Sensor Calib.


Select

“Maintenance Alarm”

by pressing

“Ok”

to select it.


You will now see:

Maintenance Alarm Activated “Yes”


Press

“Ok”

so it says

“No.”

Arrow down to

“Save and Exit.”

Press

“Ok”


Arrow down to

“Deactivation Rights”

press

“Ok”

to select it.


You will now see:

“Deactivation” “No.”


Press

“Ok”

so it says

“Yes.”

Arrow down to

“Save and Exit.”

Press

“Ok.”


Arrow down to

“Semicon S. Trig. Value”

press

“Ok”

to select it.


You will now see:

>Value
CAS number
Molecule list


Press

“OK”

to select

“Value.”


Now arrow up and down and put in

“X” for trig

. Value from your notes in step

two or refer to your GFH settings sheet. If unknown, use 3500 MV.

“Confirm new value”

and press

“Ok”

to confirm.