Secure mode can be used so that the hood operator with the “User card” cannot silence

any alarms (the sash too high) or turn the hood on/off. They can only turn the lights on/off

as it is needed during their hood operation.

Activate Secure Mode

1. Put an

“Administrator”

card in the card reader.

2. Arrow down to

“Administration Menu”

and press

“OK”

to select it.

3. Now arrow down to

“Ventilation”

and press

“OK”

to select it.

4. Next arrow down to

“Secure Mode”

and press

“OK”

to select it.

5. Now press

“OK”

to have it say

“Yes”

then arrow to

“Save and Exit”

and press

“OK”

to save the new setting.

Deactivate Secure Mode

1. Put an

“Administrator”

card in the card reader.

2. Arrow down to

“Administration Menu”

and press

“OK”

to select it.

3. Now arrow down to

“Ventilation”

and press

“OK”

to select it.

4. Next arrow down to

“Secure Mode”

and press

“OK”

to select it.

5. Now press

“OK”

to have it say

“No”

then arrow to

“Save and Exit”

and press

“OK”

to save the new setting.