Labconco Protecto Ech Filtered Fume Hoods 183 Series 8' Manuel d'utilisation

Page 100

Advertising
background image

Chapter 6: Maintaining the Protector Echo

Product Service 1-800-522-7658

94

Original instructions

Step 1 – Write down parameters under “Hood Status”


Using a maintenance card, place it in the card reader and press

“Ok”

on the

“Hood status”

menu to open it.


It may be necessary to press

“Ok”

again to open the

“Hood Status”

menu.


Then arrow up or down to record the following on paper:

DHCP: X

IP: XXX.XXX.XXX.XXX
MSK: XXX.XXX.XXX.XXX
GW: XXX.XXX.XXX.XXX


(You need this information only if you have gGuard installed.)

If DHCP is

yes

then you have a dynamic IP, if

no

then you have a static IP which

you must reenter completely.

Elapsed Time
Filter X D (this is the # of days since install)

Fans
Set point Dflt. XXXX rpm
Set point XXXX rpm

Semicon S. trig value (example 3500)
Trig. X
Delay X

Sash
Which type (example. normal polarity)
Trig X %


Now press

“X”

to return to the previous menu.”

Step 2 – Turn off the hood


Turn off the hood by pressing the fan icon on the front of the control module.
Wait 30 seconds for the hood to completely power down. If the lights are on,
press the light icon to turn the lights off as well.

Also turn off all power supplied to the unit after you have powered it down
waiting the 30 seconds.

***Very important all power supplied to hood must be turned off at the
breaker.****