Appendix a (english) – Physio-Control LUCAS Manuel d'utilisation

Page 70

Advertising
background image

70

100359-50 B, ©J

OLIFE

AB 2006

LUCAS

TM

I

NSTRUCTIONS

FOR

U

SE

G

EBRUIKSAANWIJZING

I

NSTRUCTIONS

D

UTILISATION

Appendix A (English)

W

EEKLY

CHECK

AND

CHECK

AFTER

USE

Use this protocol for the weekly check on the device and the check after use. Store the completed checklist.
Copying of this sheet is permitted.

If any error cannot be corrected take the device out of operation and indicate the problem in the “Serious problem”
column. Then send the device to the manufacturer for service immediately.

When carrying out these checks always use air from a central air supply or separate air cylinder. Do not use up
air from a air cylinder intended for use in emergency situations.

Date:

Action

Approved
signature

Adjustments
carried out

Serious problem

Check that the device is clean.

Check that the air hose is not cracked and fits
tightly to LUCAS.

Check that the connector(s) is/are present and
undamaged.

Check that the

ON/OFF

knob is in the

Adjust

position.

Pull the release rings so that the claw locks are
open for assembly. Attach the upper part to the
back plate.

Check that the compression mechanism can be
adjusted from the uppermost to the lowermost
position without any jarring or sluggishness.

Attach the air hose to a air supply. Switch the
device on for 1–2 minutes. Check that it is operating
normally. Calculate the number of compressions
per 60 seconds. 100 compressions/min ± 10 is
acceptable.